| Model |
G04h |
| Electrical requirements |
110V (±10%) 60Hz 1-Phase 680W 6A |
| Power Consumption |
5.5 kW/day |
| Performance |
- Temperature range: -40°C to -86°lC
- Time from ambient to -86°C: 4 hours
- Uniformity: ±4 °C
- Time from -86°C to -50°C with the door closed (ambient 25°C): 2 hours
|
| Controller |
- Style: microcomputer controlled with analog display
- Range -40°C to -86°C
- Step: 1°C
- Thermo sensor: PT100
- Displays: internal temperature, ambient temperature, input voltage, error code
- Indicators: Locked, running, power, alarm, door open, over temperature, battery low, condenser clean
|
| Compressor |
- Make: SECOP (formerly known as Danfoss), Germany
- Style: direct cooling
- Power: 115V 60Hz 1-PHASE, 0.54HP
- Number of compressors: two
- HC refrigerants: R600A, R1150 and R50
|
| Chamber |
- Dimensions: 13 x 19.5 x 24.5" (WxDxH)
- Material: powder coated stee
- Capacity: 3.5 Cu Ft
- Insulation: VIP (Vacuum Insulated Panel)
- 5x5x2" cryo box capacity: 60
|
| Shelf |
- Come with unit: one adjustable shelf (2 compartments)
- Material: 304 stainless steel
|
| Door and gasket |
- Inner door: two, stainless steel
- External door: one, with four layers of silicone sealing gaskets, lock-able
|
| Communications |
- Standard: RS-485, USB (stores 15 years of data for compliance purpose)
- Optional: 4-20mA transmitter
|
| Access Port |
One |
| Noise Level |
46 dBA |
| Working environment |
- Temperature: 10°C to 32°C (humidity <57% Rh required at 32°C)
- Humidity: <80% Rh
- Altitude: < 2000 m / 6561 Ft
- Stable, well-ventilated platform with no excessive dust
- Keep freezer at least 12 inches from any walls or ceiling
- Use dedicated power, do not share outlet
- Extension power cable should be less than 10 feet is required
- Freezer and power outlet should be grounded properly
- Avoid direct sunlight, water spill and do NOT use freezer outdoor
- Clean condenser fins and remove frost inside chamber at least once a month
- Important! Do not plug in your unit. Leave your Ai refrigerator or freezer for 24 hours before turning the unit on for the first time.
|
| Battery backup system |
48 hours (for controller display and alarms only)
|
Unit/shipping weight & dimensions
|
- 240 Lbs / 30 x 26 x 32"
- 300 Lbs / 33 x 29 x 35"
|
| Safety |
- Compressor over-temp protection
- Ambient over-temp protection
- Compressor power surge delay protection
- Password protection
- Sensor failure protection
- Over-temperature and low-temperature protection
- Power failure protection
- Low battery, door open, condenser clogging, system failure alarms
- External door with lock
- Ground protection
|
| Compliance |
|
| Warranty |
2-year full service, 3-year parts, 5-year compressor
|