Model |
G26h |
Electrical requirements |
110V (±10%) 60Hz 1-Phase 680W 6A |
Power Consumption |
10.5 kW/day, Energy Star Certified
|
Performance |
- Temperature range: -40°C to -86°lC
- Time from ambient to -86°C: 5 hours
- Uniformity: ±4 °C
- Time from -86°C to -50°C with the door closed (ambient 25°C): 5 hours
- 1 min door open recovery time (ambient 25°C, running temp -75°C): 30 mins
|
Dimensions |
- Top Compartment height = 11 5/8"
- Other 3 compartment height = 11 7/8"
- Width= 29.5"
- Shelf Depth= 25.5"
- Unit depth up to front support bar = 27 3/4"
|
Controller |
- Style: microcomputer controlled with analog display
- Range -40°C to -86°C
- Step: 1°C
- Thermo sensor: PT100
- Displays: internal temperature, ambient temperature, input voltage, error code
- Indicators: high/low temperature/battery alarms, sensor failure, power failure, condenser failure, door open, high ambient temperature, working condition
|
Compressor |
- Make: SECOP (formerly known as Danfoss), Germany
- Style: direct cooling
- Power: 115V 60Hz 1-PHASE, 0.85HP x 2
- Number of compressors: two
- HC refrigerants: R290 (stage 1), R170 (stage 2)
|
Chamber |
- Dimensions: 30.5 x 28.2 x 51.6" (WxDxH)
- Material: powder coated stee
- Capacity: 25.7 Cu Ft
- Insulation: VIP (Vacuum Insulated Panel)
- 5x5x2" cryo box capacity: 500
|
Shelf |
- Come with unit: 3 (4 compartments), fully adjustable
- Dimensions: 29.5 x 26.7" (WxD)
- Max. load capacity: 200 Lbs
- Material: 304 stainless steel
|
Door and gasket |
- Inner door: four, each with sealing gasket
- External door: one, with four sealing gaskets, lock-able
|
Communications |
- Standard: RS-485, USB (stores 15 years of data for compliance purpose)
- Optional: 4-20mA transmitter
|
Access Port |
Two |
Noise Level |
50 dBA |
Working environment |
- Temperature: 10°C to 32°C (humidity <57% Rh required at 32°C)
- Humidity: <80% Rh
- Altitude: < 2000 m / 6561 Ft
- Stable, well-ventilated platform with no excessive dust
- Keep freezer at least 12 inches from any walls or ceiling
- Use dedicated power, do not share outlet
- Extension power cable should be less than 10 feet is required
- Freezer and power outlet should be grounded properly
- Avoid direct sunlight, water spill and do NOT use freezer outdoor
- Clean condenser fins and remove frost inside chamber at least once a month
- Important! Do not plug in your unit. Leave your Ai refrigerator or freezer for 24 hours before turning the unit on for the first time.
|
Battery backup system |
48 hours (for controller display and alarms only)
|
Unit/shipping weight & dimensions
|
- 760 Lbs / 41 x 38.5 x 78
- 850 Lbs / 45 x 42 x 82"
|
Safety |
- Compressor over-temp protection
- Ambient over-temp protection
- Compressor power surge delay protection
- Password protection
- Sensor failure protection
- Over-temperature and low-temperature protection
- Power failure protection
- Low battery, door open, condenser clogging, system failure alarms
- External door with lock
- Ground protection
|
Compliance |
- UL 61010-1
- CSA C22.2
- Energy Star
|
Warranty |
2-year full service, 3-year parts, 5-year compressor
|